KSA-7000 Stack Gas Analyzer
Flue Gas (Stack) Analyzers

KSA-7000

Stack Gas Analyzer

Model
KSA-7000
Manufacturer
SENSORONIC
Category
Gas Analyzers › Flue Gas (Stack) Analyzers

Parameters: CO, CO2, H2S, NH3, O2, HCl, HF

Overview

KSA-7000 is a flue gas analyzer based on tunable diode laser absorption spectroscopy (TDLAS), measuring various gas components including CO, CO2, H2S, NH3, O2, HCl and HF.

It uses a hot & wet sampling method through a sampling line heated to 120~190℃, analyzing exhaust gas directly without a chiller and minimizing condensation and loss of gas components.

It provides 4~20 mA analog output for each gas and is used for coal gasification and pyrolysis, waste incinerators, cement plants and industrial stack emission measurement.

Key Features

  • Tunable diode laser absorption spectroscopy (TDLAS)
  • Measures various gases including CO, CO2, H2S, NH3, O2, HCl, HF
  • Hot & wet sampling method
  • Direct analysis of exhaust gas through a heated sampling line (120~190℃)
  • Minimized condensation and loss of gas components
  • Minimized effect of temperature and pressure fluctuations
  • High accuracy and stable analytical performance
  • 7-inch color touch-screen LCD
  • Modular design for easy maintenance
  • RS-232/RS-485 Modbus communication
  • 4~20 mA analog output for each gas
  • Hot & wet sampling without a chiller
  • Suitable for industrial sites and continuous emission monitoring systems

Measured Parameters

ParameterMeasuring Range
CO0~500mg/m³
CO20~25%vol
H2S0~200mg/m³
NH30~50mg/m³
O20~25%vol
HCl0~150mg/m³
HF0~150mg/m³

Measurement Performance

ParameterMethod · Accuracy · Resolution
COMethod: TDLAS · Accuracy: ≤±1%FS · Resolution: 0.01mg/m³
CO2Method: TDLAS · Accuracy: ≤±1%FS · Resolution: 0.01%
H2SMethod: TDLAS · Accuracy: ≤±1%FS · Resolution: 0.01mg/m³
NH3Method: TDLAS · Accuracy: ≤±1%FS · Resolution: 0.01mg/m³
O2Method: TDLAS · Accuracy: ≤±1%FS · Resolution: 0.01%
HClMethod: TDLAS · Accuracy: ≤±1%FS · Resolution: 0.01mg/m³
HFMethod: TDLAS · Accuracy: ≤±1%FS · Resolution: 0.01mg/m³

Specifications

ManufacturerSENSORONIC
Measured gasesCO, CO2, H2S, NH3, O2, HCl, HF
MethodTDLAS
Accuracy≤±1%FS
Resolution0.01mg/m³ / 0.01% (CO2, O2)
Repeatability≤1%
Response time (T90)Within 60 s
Display7-inch color touch-screen LCD
Zero drift≤2%FS/24h / ≤3%FS/7d
Span drift≤2%FS/24h / ≤3%FS/7d
Gas flow rate0.7~1.2 L/min
Sampling methodHot & wet method, no chiller required
Sampling temperature120~190℃
Analog output4~20 mA per gas
Digital communicationRS-232 / RS-485 Modbus Protocol
Power supplyAC220V / 50Hz
Dimensions400×485×177 mm (L×W×H)
Weight10kg

Applications

  • Exhaust gas analysis in coal gasification and pyrolysis
  • Gas composition analysis in plasma gasification
  • Industrial flue and stack emission measurement
  • Clean Development Mechanism (CDM) projects
  • Emission monitoring of waste incinerators and cement plants
  • Gas analysis at anaerobic digesters and wastewater treatment plants
  • Analysis of hazardous and acid gas concentrations at industrial facilities
  • Performance monitoring of pollution control and exhaust treatment equipment