KSA-1000 Flue Gas Analyzer
Flue Gas (Stack) Analyzers

KSA-1000

Flue Gas Analyzer

Model
KSA-1000
Manufacturer
SENSORONIC
Category
Gas Analyzers › Flue Gas (Stack) Analyzers

Parameters: NO, NO2, SO2

Overview

KSA-1000 is an emission gas analyzer using NDIR and non-dispersive ultraviolet (NDUV) measurement technologies, simultaneously measuring three gas components: NO, NO2 and SO2.

It reduces the effect of gas flow fluctuations and supports data logging in CSV format with a 4~20 mA output channel for each gas.

It is used for industrial stack emission monitoring, waste incineration facilities, landfill gas, and SOx and NOx measurement at industrial facilities.

Key Features

  • Simultaneous measurement of 3 gases: NO, NO2, SO2
  • NDIR and non-dispersive ultraviolet (NDUV) measurement technologies
  • Fast response (T90 within 10 s)
  • Real-time online gas concentration monitoring
  • 7-inch color touch-screen LCD
  • Automatic zero function
  • High accuracy and stable analytical performance
  • Modular sensor design for easy maintenance
  • RS-232/RS-485 Modbus communication
  • 4~20 mA analog output
  • Data logging in CSV format
  • Multiple measurement ranges available
  • Suitable for industrial sites and continuous emission monitoring systems

Measured Parameters

ParameterMeasuring Range
NO0~100/5000ppm
NO20~100/500ppm
SO20~100/5000ppm

Measurement Performance

ParameterMethod · Accuracy · Resolution
NOMethod: NDIR · Accuracy: ±2%FS · Resolution: To be confirmed
NO2Method: NDUV · Accuracy: ±2%FS · Resolution: To be confirmed
SO2Method: NDIR · Accuracy: ±2%FS · Resolution: To be confirmed

Specifications

ManufacturerSENSORONIC
Measured gasesNO, NO2, SO2
MethodNDIR / NDUV
Accuracy±2%FS
Repeatability≤1%
Response time (T90)Within 10 s
Display7-inch color touch-screen LCD
Gas flow rate0.7~1.2 L/min
Data loggingCSV Format
Gas pressure2~50 kPa
Analog output4~20 mA output channel for each gas
Digital communicationRS-232 / RS-485 Modbus Protocol
Operating temperature-15℃ ~ +45℃
Operating humidity0~95%RH (non-condensing)
Ambient pressure80~120 kPa
Power supplyAC220V / 50Hz
Dimensions400×485×132 mm (L×W×H)
Weight7.0 kg

Applications

  • Exhaust gas analysis in biomass gasification
  • Industrial stack emission monitoring
  • Gas analysis of anaerobic digesters
  • Landfill gas and exhaust gas monitoring
  • Clean Development Mechanism (CDM) projects
  • Coal gasification and pyrolysis processes
  • Plasma gasification processes
  • Biogas engines and combined heat and power (CHP) equipment
  • Exhaust gas analysis at waste incineration facilities
  • Gas measurement at industrial pollution control facilities