GasTDL-3000 Laser Ammonia Slip Gas Analyzer
Flue Gas (Stack) Analyzers

GasTDL-3000

Laser Ammonia Slip Gas Analyzer

Model
GasTDL-3000
Category
Gas Analyzers › Flue Gas (Stack) Analyzers

Parameters: NH3

Measurement PrincipleTDLAS

Overview

GasTDL-3000 is an extractive TDLAS gas analyzer with high-temperature heat tracing that prevents condensation and acid gas loss, measuring NH3, HCl or HF in real time. Multi-reflection cells give low detection limits, and the compact wall-mounted design is easy to install and maintain for flue gas treatment control and environmental monitoring.

Key Features

  • High-precision: Using specific laser wavelengths for measured gases, GasTDL-3000 can respond faster to ultra-low level gases and accurately detect specific components from gas mixtures without cross-interference.
  • Ultra-low range measurement: Using multiple reflection gas chambers, providing longer optical paths and lower detection limits.
  • Extractive method: Effectively reducing the influence of dust, temperature, and pressure fluctuations in the flue.
  • Good environmental adaptability: Suitable for high-temperature, high-humidity, strong corrosion, dusty and other harsh environment.
  • Easy installation: Compact wall-mounted design adapatble for different site installation conditions.
  • Low maintenance frequency: Easy to use and calibrate, no professional operator required and no cumbersome maintenance needed.

Measured Parameters

ParameterMeasuring Range
NH3To be confirmed

Specifications

Measurement ComponentsNH3, HCL and HF
Measurement PrincipleTDLAS
Measurement RangeNH3: / Min: 0-20ppm, Max: 0-10000ppm / HCL, HF: / Min: 0-100ppm, Max: 0-10000ppm
Resolution0.1ppm
Precision±2% F.S.
Response Time≤120s
Zero Drift≤±2%FS
Working Temperature(-10~+55) °C
Working Power Supply220V AC, 50Hz
Backflush GasClean and Dry compressed gas without oil
Warm-up Time30 min
Heating Temperature190°C
CommunicationRS485, RS232
Output Mode4-20mA, Max load 900Ω,relay output

Applications

  • Flue gas treatment system control
  • NH3, HCl and HF acid gas monitoring
  • Environmental monitoring