Gasboard-2601 GFC Infrared Gas Sensor Bench
OEM Gas Benches & Analysis Modules

Gasboard-2601

GFC Infrared Gas Sensor Bench

Model
Gasboard-2601
Category
Gas Analyzers › OEM Gas Benches & Analysis Modules

Parameters: CO, CO2, CH4, N2O

Measurement PrincipleGFC

Overview

Gasboard-2601 is a gas filter correlation (GFC) infrared sensor bench that compares sample and reference absorption to eliminate interference, detecting CO, CO2, CH4 or N2O below 0-50 ppm. TEC cooling with PID temperature control minimizes drift, and its modular design and RS-232 calibration ease integration.

Key Features

  • Good anti-interference performance based on Gas filtering correlation (GFC) infrared absorption technology and narrow band filtering structure.
  • Suitable for ultra-low concentration gas measurement, can detect CO/CO2/CH4/N2O gas below 0-50ppm.
  • Small drift, high stability, less affected by the external environment, no need for frequent calibration by using TEC refrigeration, self-setting PID constant temperature control.
  • Meet the long-term stability of work with Long life light source and reliable filter gas sealing technology.
  • Independent gas detection modular design, rich interface, easy to expand and integrate.

Measured Parameters

ParameterMeasuring Range
COTo be confirmed
CO2To be confirmed
CH4To be confirmed
N2OTo be confirmed

Specifications

Measurement GasCO, CO2, CH4, N2O
Range/ResolutionCO: 0-50ppm, 0.01ppm / CO2: 0-100ppm, 0.01ppm / CH4: 0-50ppm, 0.01ppm / N2O: 0-20ppm, 0.01ppm / (Measurement range can be customized)
Linear Error≤±2%F.S.
Response Time≤60s@T90
Repeatability≤1%
Zero Drift±0.2ppm/24h
Span Drift±0.5ppm/24h
Warm-up Time45min
Ambient Temperature(0~45)℃
Relative Humidity0%~90%rel (non-condensing)
Inlet Pressure850mBar ~1100mBar
Dust< 0.5μm
CommunicationRS232
Power Supply12VDC±0.2V/4A / 220VDC±10%
Power Consumption<150W

Applications

  • Ambient air monitoring
  • Low-level greenhouse gas measurement
  • OEM analyzer integration