Gasboard-2006 Optical NDIR Multi-Gas Sensor
OEM Gas Benches & Analysis Modules

Gasboard-2006

Optical NDIR Multi-Gas Sensor

Model
Gasboard-2006
Category
Gas Analyzers › OEM Gas Benches & Analysis Modules

Parameters: Multi-gas

Measurement PrincipleNDIR

Overview

Gasboard-2006 is an NDIR multi-channel sensor bench that measures CO and CO2 (optional CH4) simultaneously at constant temperature without cross-interference. With CO ranges down to 0-1000 ppm, linearity error below 1% F.S. and interference-compensating algorithms, it integrates easily into multi-gas systems.

Key Features

  • Measure CO, CO2 (and CH4) simultaneously in one gas sensor bench.
  • Low CO measurement range down to 0-1000ppm.
  • Excellent performances and high accuracy with linear error less than 1% F.S.
  • Constant temperature design ensures high precision and stability by reducing the influence from ambient temperature.
  • Advanced mathematical algorithm identifies optimal gas absorption peak and compensates for interference from other gases, improving measurement accuracy.
  • Modular design for easy integration.

Measured Parameters

ParameterMeasuring Range
Multi-gasTo be confirmed

Specifications

Measurement GasCO, CO2 (CH4 optional on request)
Range/ResolutionCO: 0-1000ppm, 0.1ppm / CO2: 0-20% Vol., 0.01% Vol. / * CH4 measurement on request
LODCO: 0.5%F.S. / CO2: 0.1%F.S.
Linear Error≤±1%F.S.
Response Time≤8s@0.7L/min
Repeatability≤1%
Stability≤±2%F.S./24h
Constant Temperature50℃
Warm-up Time<30min
Ambient Temperature(5~40)℃
Relative Humidity0%~95%rel (non-condensing)
Inlet Pressure(2~50)kPa
Flow Rate0.2L/min~1L/min
CommunicationRS485/RS232 Modbus RTU
Power Supply12VDC±0.2V/4A
Power Consumption<800mW/Max. 36W during heating

Applications

  • Multi-gas analyzer integration
  • Combustion and process gas measurement