Gasboard-2000L NDIR Gas Bench 2000L
OEM Gas Benches & Analysis Modules

Gasboard-2000L

NDIR Gas Bench 2000L

Model
Gasboard-2000L
Category
Gas Analyzers › OEM Gas Benches & Analysis Modules

Parameters: HC, CO, CO2

IndustriesAutomotive
Measurement PrincipleNDIR

Overview

Gasboard-2000L is a compact NDIR gas bench for CO, CO2 and HC in gasoline and diesel exhaust, expandable with modular O2 and NOx sensors. OIML R99 Class 0 compliant, it adds automatic temperature compensation, integrated differential and absolute pressure sensors for leak testing and a fast warm-up.

Key Features

  • Integrated NDIR gas bench for CO, CO2, and HC measurement.
  • Automatic temperature compensation for accurate and stable measurement.
  • Integrated differential and absolute pressure sensors for leak testing and pressure compensation.
  • Fast warm-up time for quick system readiness.
  • Modular design for customized solutions.

Measured Parameters

ParameterMeasuring Range
HCTo be confirmed
COTo be confirmed
CO2To be confirmed

Specifications

Detecting PrincipleNDIR (non-dispersive infrared) technology
MeasurementHC (n-Hexane) | CO | CO2
Range (vol)0-20000 (×10-6) | 0-15% | 0-20%
Resolution (vol)1×10-6 | 0.00% | 0.01%
Range (vol)0-2000 | 2001-5000 | 5001-20000 | 0-10% | 10.01%-15% | 0-16% | 16.01%-20%
(×10-6)(×10-6)
Abs. Error±10 (×10-6) | ±0.03% | ±0.4%
Rel. Error±5% | ±10% | ±3% | ±4%
Response Time (T90)≤4s
Gas Flow0.7-1.2L/min
Warm-up Time1 minute usable, 3 minutes to full performance
Power Supply12V / 0.15A
Power Consumption2W
Ambient Temperature(5~45) °C
Relative Humidity0~90% RH (No condense water)
Atmospheric Pressure(75~110) kPa
CommunicationRS232

Applications

  • Emission analyzer manufacturing
  • OEM system integration