Biogas Analyzers
KSA-2100UV
UV-DOAS Biogas Analyzer
- Model
- KSA-2100UV
- Manufacturer
- SENSORONIC
- Category
- Gas Analyzers › Biogas Analyzers
Parameters: CH4, CO2, H2S, O2
IndustriesEnergy & Biogas
Measurement PrincipleUV-DOAS
Overview
KSA-2100UV is a biogas analyzer using UV absorption spectroscopy (UV-DOAS) and NDIR sensor technology to measure the main biogas components CH4, CO2, H2S and O2.
UV-DOAS technology provides stable analysis even at high H2S concentrations, and the design minimizes biogas loss during measurement.
It supports a 7-inch touch screen, real-time gas concentration monitoring and RS-232/RS-485 communication, and is used for process control and quality control at anaerobic digesters, landfill gas sites and biogas plants.
Key Features
- Simultaneous measurement of 4 gases: CH4, CO2, H2S, O2
- High-concentration H2S measurement using UV-DOAS
- NDIR and electrochemical (ECD) sensors
- High accuracy and reliability with UV-DOAS technology
- Stable gas analysis even at high H2S concentrations
- Designed to minimize biogas loss during measurement
- 7-inch touch-screen LCD display
- Real-time gas concentration monitoring
- 4~20 mA analog output
- RS-232/RS-485 communication
- Continuous monitoring for optimized process control
- Modular design for easy calibration
- Automatic zero function
- Analysis system suited to biogas production and treatment facilities
Measured Parameters
| Parameter | Measuring Range |
|---|---|
| CH4 | 0~100%vol |
| CO2 | 0~50%vol |
| H2S | 0~500/5000ppm |
| O2 | 0~25%vol |
Measurement Performance
| Parameter | Method · Accuracy · Resolution |
|---|---|
| CH4 | Method: NDIR · Accuracy: ±2%FS · Resolution: 0.01 |
| CO2 | Method: NDIR · Accuracy: ±2%FS · Resolution: 0.01 |
| H2S | Method: UV-DOAS · Accuracy: ±2%FS · Resolution: 1 |
| O2 | Method: ECD · Accuracy: ±3%FS · Resolution: 0.01 |
Specifications
| Manufacturer | SENSORONIC |
|---|---|
| Measured gases | CH4, CO2, H2S, O2 |
| Method | NDIR / UV-DOAS / ECD |
| Accuracy (CH4, CO2, H2S) | ±2%FS |
| Accuracy (O2) | ±3%FS |
| Repeatability | ≤2% |
| Response time (T90) | Within 60 s |
| Display | 7-inch touch-screen LCD |
| Gas flow rate | 0.7~1.2 L/min |
| Gas pressure | 2~50 kPa |
| Analog output | 4~20mA |
| Communication | RS-232 / RS-485 |
| Automatic zero function | Supported |
| Operating temperature | -15℃ ~ +45℃ |
| Operating humidity | 0~95%RH |
| Ambient pressure | 80~120 kPa |
| Power supply | AC220V, 50Hz |
| Dimensions | 132×485×332 mm (H×W×D) |
| Weight | 18 kg |
Applications
- Biogas composition analysis of anaerobic digesters
- Biogas engines and combined heat and power (CHP) plants
- Landfill gas monitoring
- Process control and quality control at biogas plants
- Hydrogen sulfide measurement at H2S scrubbers
- Upgrading and biomethane production facilities
- Biogas composition analysis at wastewater treatment plants
- Continuous gas monitoring at biogas production facilities
Related Products

Biogas Analyzers
KSA-2000
Biogas Analyzer
Parameters: CH4, CO2, H2S, O2, Flow

Biogas Analyzers
KSA-2000P
Portable Biogas Analyzer
Parameters: CH4, CO2, H2S, O2, H2

Biogas Analyzers
KSA-2000PLUS
Portable Biogas Analyzer
Parameters: CH4, CO2, H2S, O2, CO, H2

Biogas Analyzers
KSA-2300H
UV-DOAS Hydrogen Sulfide (H₂S) Analyzer
Parameters: H2S