Gasboard-9021 (Tri-Gas) Online Tri-Gas Analysis System (CO/CO₂/CH₄)
Process Gas Analyzers

Gasboard-9021 (Tri-Gas)

Online Tri-Gas Analysis System (CO/CO₂/CH₄)

Model
Gasboard-9021 (Tri-Gas)
Category
Gas Analyzers › Process Gas Analyzers

Parameters: CO, CO2, CH4 (optional: H2)

Measurement PrincipleNDIR

Overview

The Gasboard-9021 tri-gas system measures CO, CO2 and CH4 in heat treatment furnace atmospheres in real time with temperature-controlled NDIR sensors. It supports furnace diagnosis, carbon potential calculation and cross-checking of oxygen probes, meets CQI-9 audit requirements and offers an optional TCD module for H2.

Key Features

  • Stable NDIR Tri-Gas Detection: Temperature-controlled NDIR sensors for simultaneous measurement of CO, CO2, and CH4 with minimized drift during long-term operation.
  • CQI-9 Compliance: Meets automotive supply chain audit requirements for process and quality stability.
  • Easy Operation & Maintenance: Integrated gas sampling and conditioning system with modular design for simplified installation and maintenance.
  • Optional H2 Measurement: Optional TCD detector with advanced interference correction technology for accurate H2 measurement without cross-interference from CO, CO2, or CH4.

Measured Parameters

ParameterMeasuring Range
COTo be confirmed
CO2To be confirmed
CH4To be confirmed
H2 (optional)To be confirmed

Specifications

Measuring ComponentsCO, CO2, CH4
Measuring PrincipleNDIR
Measurement RangeCO2:0-5000ppm; / CO:0-30%; / CH4:0-10%;(Range can be customized)
Accuracy≤±2%F.S.
Repeatability≤±1%
Response Time<30S
Drift≤±2%F.S. /2 month
ResolutionCO2: 0.1ppm; CO: 0.01%; CH4: 0.01%
Inlet Flow Rate0.7~1.2L/min
Inlet Gas Pressure2~50kPa
Sampling Gas ConditionNo dust, no vapor, no tar
Operating PowerAC110/AC220V±40V, 50/60Hz
Communication InterfaceRS-485/RS-232
Output Mode4~20mA Output

Applications

  • Heat treatment furnace atmosphere monitoring
  • Carbon potential calculation
  • CQI-9 heat treatment audits